Agilent Technologies E6380-90073 Cell Phone Accessories User Manual


 
148 Chapter 1
GPIB Command Dictionary
CGENerator subsystem
:EBNO:LEVel <real number>
:EBNO:LEVel?
These commands set/query the level of the generator when it is in
Eb/No mode. This command utilizes the “Real Number Setting Syntax”
on page 299, but does not use the :STATe command.
Syntax
CGEN:EBNO:LEV <real number> !values from 5 to +25 dB
CGEN:EBNO:LEV?
Screen/field equivalent
:EBNO:LEV controls the generator’s level when the Gen Mode field on
the CDMA GENERATOR screen is set to Eb/No.
:EQFilter <string>
:EQFilter?
These commands set/query the setting of the equalization filter used to
create forward channel signals. This command is used when you want
to override the default setting (EQ In). The filter is bypassed for reverse
link signals and cannot be overridden. Therefore, this command is valid
only when the CGEN:DIR ‘Fwd’ command is used.
Syntax
CGEN:EQF ‘EQ In’ !switches the Equalization filter in
‘Out’ !bypasses the Equalization filter
CGEN:EQF?
Screen/field equivalent
CGEN:EQF controls the EQ In/Out subfield of the Gen Dir field on the
CDMA GENERATOR screen. This field is displayed only when Fwd is