Agilent Technologies E6380-90073 Cell Phone Accessories User Manual


 
258 Chapter 1
GPIB Command Dictionary
MEASure subsystem
:CDANalyzer:IS2000:TRACe:POWer:POWer:ACTive?
IS-2000 only. This query returns the active channel information for
theCode Domain trace when making a Power and Noise measurement.
The return value consists of an integer representing the number of
active channels followed by a four value record for each channel. The
four values for each record are channel number, spread factor for this
channel, Power for this channel and Noise for this channel.
Syntax
MEAS:CDAN:IS2000:TRAC:POW:POW:ACT?
Screen/field equivalent
There are no fields on the CODE DOMAIN screen equivalent to the
return value for this query. The return information is only accessible
when the CDMA standard is set to “IS-2000”, the measurement selected
is Power and Noise, and CD pwr unit is set to Rel on the Reference
image of the CODE DOMAIN screen.
:CDANalyzer:IS2000:TRACe:POWer:POWer:ALL?
IS-2000 only. This query returns the information necessary to recreate
the Code Domain trace when making a Power and Noise measurement.
The return consists of 128 four value records for each bar in the trace.
The four values for each record are channel number, Power for this
channel, Noise for this channel a flag indicating that the next bar
should be connected to this bar as part of a wider supplemental
channel.
If the selected location is part of a supplemental channel, all the Walsh
channels that comprise the supplemental channel will be returned total
power and noise value for the supplemental channel will be returned.
Syntax
MEAS:CDAN:IS2000:TRAC:POW:POW:ALL?
Screen/field equivalent
There are no fields on the CODE DOMAIN screen equivalent to the
return value for this query. The return information is only accessible
when the CDMA standard is set to “IS-2000”, the measurement selected
is Power and Noise, and CD pwr unit is set to Rel on the Reference
image of the CODE DOMAIN screen.