Agilent Technologies E6380-90073 Cell Phone Accessories User Manual


 
Chapter 1 255
GPIB Command Dictionary
MEASure subsystem
:CDANalyzer:IS2000:TRACe:POWer:COMPlex:I[:VALue]?
:CDANalyzer:IS2000:TRACe:POWer:COMPlex:COMBined[:VALue]?
IS-2000 only. This query returns an array of 128 values that represents
the Spreader I Input Power of both IS-95 and IS-2000 channels relative
to the selected reference power value.
NOTE
The values returned represent the absolute power of individual Walsh
channels not the supplemental channels of which they may be a part.
The power of supplemental channels must be read directly from
I Pwr Level fields on the CODE DOMAIN screen when Complex
Power measurement has been selected.
Syntax
MEAS:CDAN:IS2000:TRAC:POW:COMP:I[:VAL]?
MEAS:CDAN:IS2000:TRAC:POW:COMP:COMB[:VAL]?
Screen/field equivalent
MEAS:CDAN:IS2000:TRAC:POW:COMP:I[:VAL]? and
MEAS:CDAN:IS2000:TRAC:POW:COMP:COMB[:VAL]? returns
information (top set of bars) displayed for the Complex Power
measurement when CD pwr unit is set to Rel on the Reference
menu on the CODE DOMAIN screen when the CDMA standard is set to
“IS-2000”.