Agilent Technologies E6380-90073 Cell Phone Accessories User Manual


 
126 Chapter 1
GPIB Command Dictionary
CDANalyzer subsystem
:PNINcrement <real number>
:PNINcrement?
These commands set/query the PN increment. If you know the PN
incrment, enter it with this command to speed the PN offset search. If
you do not know the increment, enter a 1.
This command utilizes the :INCR command of the “Real Number
Setting Syntax” on page 299.
Syntax
CDAN:PNIN <real number>!values from 1 to 256
CDAN:PNIN? !returns the value of the PN increment
Screen/field equivalent
CDAN:PNIN controls the PN Incrment field on the CODE DOM screen
when the CDMA standard is set to “IS-95 Only”. This field is displayed
only when the Find PN field on the PN Setup menu is set to Auto.
:PNMode <string>
:PNMode?
These commands set/query the PN offset mode.
‘Auto’ is used to calculate the system’s PN offset if it is unknown.
When ‘Auto’ is selected, the CDAN:PNIN command can be used to
enter a PN increment and speed the PN offset calculation.
‘Manual’ is used to enter a known PN offset. When ‘Manual’ is
selected, the PN offset is entered using the CDAN:PNOF command.
Syntax
CDAN:PNM ‘Auto’
‘Manual’
CDAN:PNM? !returns the PN Offset mode
Screen/field equivalent
CDAN:PNM controls the Find PN field on the PN Setup menu of the
CODE DOM screen, when the CDMA standard is set to “IS-95 Only”.