MEASure subsystem
:CDANalyzer:TIME:OFFSet <meas cmnd>
:CDANalyzer:TIME:OFFSet?
IS-95 Only mode. These commands set/query the time offset
measurement of the code domain analyzer.
This measurement utilizes the Number Measurement Syntax, but does
not use the :METer command.
Syntax
MEAS:CDAN:TIME:OFFS <meas cmd>
!See “Number Measurement Syntax” on page 283.
MEAS:CDAN:TIME:OFFS?
Example
MEAS:CDAN:TIME:OFFS? !displays the measurement
Screen/field equivalent
MEAS:CDAN:TIME:OFFS? reads the Time Ofs measurement field on
the CODE DOM screen when the CDMA standard is set to “IS-95
Only”.
:CDANalyzer:TRACe:POWer?
IS-95 Only mode. This command queries the power in each of the 64
Walsh channels, relative to the pilot’s power. (Pilot power is
approximately two-thirds of the total power.) The measurement returns
64 values (in dB), one for each point.
Syntax
MEAS:CDAN:TRAC:POW?
Screen/field equivalent
MEAS:CDAN:TRAC:POW? does not correspond to any field on the
CODE DOM screen. This applies when the CDMA standard is set to
“IS-95 Only”.