240 Chapter 1
GPIB Command Dictionary
MEASure subsystem
:CDANalyzer:IS2000:MARKer:NOISe?
IS-2000 only. This query returns the Marker Level for the Noise, when
the Power & Noise measurement is active. The return will be a floating
point number.
Syntax
MEAS:CDAN:IS2000:MARK:NOIS?
Screen/field equivalent
MEAS:CDAN:IS2000:MARK:NOIS? reads the Noise Lvl field in the
RH column of the Marker menu on the CODE DOMAIN screen when
the CDMA standard is set to “IS-2000” and measurement is Power &
Noise.
:CDANalyzer:IS2000:MARKer[:POWer]?
IS-2000 only. This query returns the Marker Level, in dB, when the
Channel Power Mode is inactive. The return will be a floating point
number.
Syntax
MEAS:CDAN:IS2000:MARK[:POW]?
Screen/field equivalent
MEAS:CDAN:IS2000:MARK[POW]? reads the Marker Lvl field of the
CODE DOMAIN screen when the CDMA standard is set to “IS-2000”
and the CD pwr unit field on the Reference menu is set to Rel.
:CDANalyzer:IS2000:PNumber:OFFSet?
:CDANalyzer:IS2000:PNUMber:OFFSet?
IS-2000 only. This query returns the measured PN offset of the Pilot
signal as an integer value.
Syntax
MEAS:CDAN:IS2000:PN:OFFS?
MEAS:CDAN:IS2000:PNUM:OFFS?
Screen/field equivalent
MEAS:CDAN:IS2000:PN:OFFS? and
MEAS:CDAN:IS2000:PNUM:OFFS? queries the PN Ofs measurement
field on the CODE DOMAIN screen when the CDMA standard is set to
“IS-2000”.