Agilent Technologies E6380-90073 Cell Phone Accessories User Manual


 
164 Chapter 1
GPIB Command Dictionary
CONFigure subsystem
:RFDisplay <string>
:RFDisplay?
These commands set/query the RF display mode. See the Reference
Guide for more information about frequency and channel tuning.
Syntax
CONF:RFD ‘Freq’ !sets RF display to frequency tuning mode
‘Chan’ !sets RF display to channel tuning mode
CONF:RFD?
Screen/field equivalent
CONF:RFD controls the RF Display field of the INSTRUMENT
CONFIGURE screen.
:RFIMped <string>
:RFIMped?
These commands set/query the way that RF generator’s voltages are
expressed (across a 50 ohm load or open circuit). The RF generator’s
amplitude units must be V, mV, uV, or dBuV for this command to have
an effect.
Syntax
CONF:RFIM ‘50 ohm’
‘emf’
CONF:RFIM?
Screen/field equivalent
CONF:RFIM controls the RF Gen Volts field of the INSTRUMENT
CONFIGURE screen.