Agilent Technologies E6380-90073 Cell Phone Accessories User Manual


 
Chapter 1 211
GPIB Command Dictionary
MEASure subsystem
MEASure subsystem
The MEAS subsystem has a set of unique commands. These are the
measurement command <meas cmnd> set. They control features of the
Test Set such as setting measurement limits, units and the meters. See
“Number Measurement Syntax” on page 283 for more details about
these commands.
Most commands have both a <meas cmnd> parameter and a query
command.
The command using <meas cmnd> allows you to set certain
parameters of the measurement such as limits, averages, and units.
The query command is the command that reads the measurement
and returns a value.
:RESet
This command resets all measurements in progress.
Syntax
MEAS:RESET
Screen/field equivalent
MEAS:RESET does not have an equivalent control field on the Test Set.
:AFRequency:ACLevel <meas cmnd>
:AFRequency:ACLevel?
These commands set/query the AC level measurement. It measures the
AC level of the audio source (SSB Demod, Audio In, Ext Mod, or Audio
Out) as selected by the AFAN:INP command on page 59.
Syntax
MEAS:AFR:ACL <meas cmd>
!See “Number Measurement Syntax” on page 283.
MEAS:AFR:ACL? !returns real value
Screen/field equivalent
MEAS:AFR:ACL? reads the AC Level measurement field on the AF
ANALYZER screen when the AF Anl In field is set to Audio In, Audio
Out, Ext Mod, or SSB Demod. This measurement is also displayed on the
RF ANALYZER and RF GENERATOR screens.