Agilent Technologies E6380-90073 Cell Phone Accessories User Manual


 
312 Chapter 1
GPIB Command Dictionary
RFGenerator subsystem
:MODulation:EXTernal:FM <real number>
:MODulation:EXTernal:FM?
These commands set/query the FM deviation of the RF generator when
FM is applied through the modulation input port. The modulation input
must be set to FM.
This command utilizes the “Real Number Setting Syntax” on page 299.
Syntax
:MOD:EXT:FM <real number>
:MOD:EXT:FM?
Screen/field equivalent
RFG:MOD:EXT:FM controls the lower subfield of the Mod In To field
on the RF GENERATOR screen when FM (/Vpk) is selected in the
upper subfield.
:OUTPut <string>
:OUTPut?
These commands set/query the RF output port.
Syntax
RFG:OUTP ‘RF Out’
‘Dupl’
RFG:OUTP?
Screen/field equivalent
RFG:OUTP controls the Output Port field on the RF GENERATOR
screen. This field is also displayed on the spectrum analyzer’s RF Gen
menu when the spectrum analyzer is in fixed mode.