3-25
Verifying Performance
Agilent Technologies 8922M/S Specifications
Audio Source
Specifications
Frequency
Output Level
Audio Analyzer
Specifications
Frequency Measurement
Range DC to 25 kHz
Accuracy 0.025% of setting
Supplemental Characteristics
Minimum Resolution
0.1 Hz
Range 0.1 mV to 4 V
rms
Maximum Output Current 20 mA peak
Output Impedance <1 Ω
Accuracy
±(2% of setting + resolution)
Residual Distortion (THD + noise,
amplitude >200 mV
rms
)
0.1%, 20 Hz to 25 kHz in 80 kHz BW.
Supplemental Characteristics
Minimum Resolution Level
≤0.01V: 50µV
Level
≤0.1V: 0.5 mV
Level
≤1V: 5 mV
Level >1V: 50 mV
DC Coupled Offset
<50 mV
Range 20 Hz to 400 kHz
Accuracy
±(0.02% + 1 count + reference accuracy)
External Input 20 mV
rms
to 30 V
rms
Supplemental Characteristics
Minimum Resolution
f < 10kHz: 0.01 Hz
f < 100kHz: 0.1 Hz
f
≥ 100kHz: 1 Hz