Agilent Technologies 8922M Cell Phone User Manual


 
8-32
Instrument BASIC
Programming and Using the TESTS Subsystem
Using Autostart
Ensure the Autostart field toggle is set to On (see Figure 8-13 on page 8-32 (1));
this allows the Agilent 8922M/S to go straight to the Procedure Menu each time the
Agilent 8922M/S is switched on, providing a Memory Card is inserted in the front
panel of the Agilent 8922M/S.
If the Procedure Menu screen does not appear on the Agilent 8922M/S display,
select and load the procedure.
Figure 8-13 Tests Screen of the Agilent 8922M/S
Making a Test Sequence
The Test Sequence menu lets you select radio tests and the order you want them
performed.
1 Select the Edit Seqn field. A test-sequence screen similar to that shown in
Figure 8-14 on page 8-33 appears.
2 Select the Insrt Stp field or Delet Stp field as needed to insert or delete tests on
the screen. Refer to item (1) in Figure 8-14 on page 8-33.
3 Select the Step # field and turn the knob to the test you want to change. (As
you turn the knob, the test’s step number, name, and description changes.
4 Select the Test Name field and turn the knob to select the desired test. Refer to
item (4) in Figure 8-14 on page 8-33. (As you turn the knob, the test name and
description changes along with the test number shown in the Choices menu.
TESTS