4-5
Screens
Audio
Choices Scope In selects the SCOPE IN (MEASURE) front-panel
connector.
Speech Out selects the signal going to the MON/SPEECH
(DEMODULATION OUT) front-panel connector.
AM Mod In selects the IN AM/SPEECH (MODULATION) front-
panel connector.
Speech In selects the IN AM/SPEECH (MODULATION) front-
panel connector.
FM Demod selects the FM demodulation discriminator.
Pls Demod selects the pulse demodulation detector.
Audio In selects the (AUDIO IN HI) front-panel connector.
Audio Out selects the signal going to the AUDIO OUT front-
panel connector.
NOTE The AM Mod In and the Speech In are not featured in the Agilent 8922S.
4. Amplitude This field is the AF Generator Audio Out amplitude setting.
Range 0.0mV rms to 8.84 V rms.
5. Audio In Lo This field sets the reference for the Audio In Lo BNC connector. The choices are
Float or Gnd.
6. Coupling This field selects ac or dc coupling of the AF Generator Audio Out signal.
7. Echo Delay This field sets the echo delay when the Speech field is set to Echo mode.
8. Ext Load R This field sets the AF analyzer external load resistance for displaying AF analyzer
measurement results in Watts.
Range 0.1 Ω to 1 MΩ.
9. Frequency This field is the AF Generator Audio Out Frequency setting.
Range dc to 30 kHz.