Agilent Technologies 8922M Cell Phone User Manual


 
4-89
Screens
Phase and Frequency Error - Multi-burst ON
3. Errors This field displays the number of bursts which could not be used as part of an overall
calculation. This field is updated when the measurement is completed. When the
measurement cycle is complete, this field will display the total number of unused
bursts.
See Also Screens: Phase and Frequency Error; Wanted, Done
4. Meas Frequency This field displays the frequency at which measurements are being made. This field
is only used in the TEST MODE.
5. ON/OFF This field enables or disables multi-burst measurement. The default for this field is
OFF.
Choices ON enables multi-burst measurement.
OFF disables multi-burst measurement.
6. Peak Phase
RMS
Frequency Error
These fields show the measurement results made during a measurement cycle when
multi-burst is on. The table below explains each of the measurements made.
7. Single/Cont Single allows one test to be performed. Cont allows testing to automatically repeat.
Error Mean Maximum Minimum Last
Peak phase Mean of the
peak phase
errors of
each burst
Largest peak
phase error
in any of the
bursts
Smallest
peak phase
error in any
of the bursts
Peak phase
error of the
last burst
RMS phase Mean of the
RMS phase
errors of each
burst
Largest RMS
phase error
in any of the
bursts
Smallest
RMS phase
error in any
of the bursts
RMS phase
error of the
last burst
Frequency Mean of the
frequency
errors of each
burst
MostPositive
frequency
error in any
of the bursts
Most
Negative
frequency
error in any
of the bursts
Frequency
error of the
last burst