Agilent Technologies 8922M Cell Phone User Manual


 
3-26
Verifying Performance
Agilent Technologies 8922M/S Specifications
AC Voltage Measurement
DC Voltage Measurement
Distortion Measurement
Voltage Range 0 V to 30 V
rms
Accuracy (20 Hz to 15 kHz),
Input >1 mV
rms
±3% of reading
Residual Noise + THD (15 kHz BW) 175 µV
Supplemental Characteristics
3 dB Bandwidth
2 Hz to 100 kHz
Input Impedance
1 M 145 pF at AUDIO IN
Minimum Resolution 4 digits for inputs
100mV
3 digits for inputs <100 mV
Voltage Range 100 mV to 42 V
Accuracy
±(1.0% of reading + DC Offset)
DC Offset:
±45 mV
Supplemental Characteristics
Minimum Resolution
1.0 mV
Fundamental Frequency
1 kHz
±5 Hz
Input Level Range 30 mV
rms
to 30 V
rms
Display Range 0.1% to 100%
Accuracy
±1 dB (0.5 to 100% distortion)
Residual THD + noise (15 kHz BW)
The greater of
60 dB or +175 µV
Supplemental Characteristics
Minimum Resolution
0.01% distortion