Instrument BASIC
Programming and Using the TESTS Subsystem
To Select Printing Conditions
Refer to item (3) in Figure 8-12 on page 8-30.
Output Results.….……All
All test results are shown on the output device (CRT and/or printer). Printouts
include a “banner” listing the test conditions, measured values, lower and upper
limits, and whether the test passed or failed. The Comment field is shown at the top
along with any identifying information from the Output Heading field. Date,
and time is also output.
Output Results.….……Failures
Test results are shown only when a UUT failure or software error occurs. Printouts
include a “banner” listing the test conditions, measured values, and lower and upper
limits of the failed test. The Comment field and any identifying information from
the Output Heading field is also output.
To Have Test Results Appear on a CRT or Printer
Refer to item (4) in Figure 8-12 on page 8-30.
Output Destination.….……Crt
Test results are output to the Agilent 8922M/S CRT screen only.
Output Destination.….……Printer
Test results are output to the CRT and printer. A printer must be correctly configured
in order to get a printout.
To configure an RS-232 printer, refer to chapter 4 for I/O CONFIGURE screen
descriptions.
To configure a GPIB printer, refer to the instructions in the following section
titled Configuring External Instruments for GPIB Control.
To Enter Comments in the Output Heading Field
Refer to item (5) in Figure 8-12 on page 8-30.
1 Select the Output Heading field. (An alpha/numeric list of characters
appears in the lower-right corner of the screen.)
2 Select characters one at a time using the knob in order to compose the comment
you want to make. (Two lines of comments, 50 characters in length, may be
entered.)
3 Select Done when you are finished.