Agilent Technologies 8922M Cell Phone User Manual


 
4-12
Screens
Bit Error 2
NOTE This field is not featured in the Agilent 8922S.
2. Amplitude This field changes the RF generator amplitude. It is a duplicate of the RF generator’s
Amplitude field on the RF Generator/RF Analyzer screen.
See Also Screens: RF Generator/RF Analyzer (RF Generator).
3. Atten Hold This field turns the attenuator hold on or off.
4/5. BE Ratio/BE
Count
These fields select how bit errors will be displayed for both the Intermediate and
Completed results.
Choices Bit Error displays the bit errors/count per measurement.
Bit Count calculates and displays the ratio of bit errors counted to
the total number of bits measured (depends on measurement type).
Field (4) displays the final result of the measurement, field (5) displays a running
total of the bit errors.
6. Meas Num Use this field to define and display measurements 1 or 3. The right side of the screen
allows access to measurements 2 or 4. Refer to 16 for further information.
7. Bits Tested
(Intermediate)
This field displays the number of bits that have been tested during a measurement
that is currently running. This number includes only bits that are tested by the
selected measurement type.
8. Bits Tested
(Completed)
This field displays the number of bits that were tested during the last fully completed
test.
Bit error tests may extend beyond the number entered in the “Bits To Test” field so
that the last speech frame may be completed.
9. Bits To Test This field selects the number of bits which will be tested. This number includes only
bits that are tested by the selected measurement type.
10. CRC or FE These fields determine what data will be displayed.
Choices CRC displays cyclic-redundancy-check (CRC) errors.
FE displays frame erasures.
See Also Keys: ON/OFF, LO LIMIT, HI LIMIT, REF SET, AVG, Units