Agilent Technologies 8922M Cell Phone User Manual


 
4-55
Screens
CW Measurement
3. CW Power This field displays the measured CW power. This measurement is valid only for
non-pulsed signals at the front-panel RF IN/OUT connector. This measurement is
made at the frequency entered in the Frequency field.
See Also Keys: ON/OFF, HI LIMIT, LO LIMIT, REF SET, AVG,
METER, Units
4. Frequency
(Expected Input)
This field sets the frequency of the signal to be measured. It needs to be within ±500
kHz to ensure the correct readings of amplitude (3) and frequency (2).
In the GSM band, the value that is entered should be set to the nearest 100 kHz.
5. Meas Reset (Meas
Cntl)
Selecting Meas Reset will erase any accumulated measurements used for calculating
a final result, and re-start the measurement process for the following functions:
HI LIMIT
LO LIMIT
AVG
This field resets the AF Analyzer measurements.
See Also Keys: HIGH LIMIT, LO LIMIT, AVG
6. Pwr Zero (Meas
Cntl)
This field zeros the power meter. RF power must be disconnected from the RF
IN/OUT port when executing this function.
7. RF Cnt Gate This field sets the frequency counter’s gate time.
Range 0.11 seconds to 1 second.
8. RF Input This field selects the front-panel RF input for the RF analyzer.
Choices RF IN/OUT
AUX RF IN