Agilent Technologies 8922M Cell Phone User Manual


 
6-16
Connectors
Rear-Panel Connectors of the Agilent Technologies 8922M/S
9. Protocol Interface This port is only available if Option 003 is fitted. It allows protocol messages to be
sent to an external protocol logger.
Refer to Appendix A for more information.
10. REF IN The timebase reference input is a BNC connector. It is connected to the reference
circuitry. This connector is the main timebase reference input. Normally, it is
connected to Option 001 REF OUT (if option 001 is installed) using an Agilent-
supplied short jumper cable. RF IN can also be connected to a user-provided 1, 2, 5,
10 or 13 MHz reference by selecting the appropriate frequency from the Reference
field on the Configure screen.
If RF IN is left unconnected, the internal timebase will not be locked to any external
reference.
See Also
Screens: Configure Specifications
11. SCOPE This connector is used in conjunction with the HP/Agilent 83220A and HP/Agilent
83220E DCS/PCS Test Sets. Refer to either the HP/Agilent 83220A or HP/Agilent
83220E User’s Guides for further information.