Agilent Technologies 8922M Cell Phone User Manual


 
4-54
Screens
CW Measurement
CW Measurement
1. Amplitude
(Expected Input)
This field is the input amplitude to be assumed at the selected RF analyzer input.
The amplitude shown is for the port selected in the RF Input field.
Range RF IN/OUT: 27.9 to +41.0 dBm. AUX RF IN: 58.0 to +20.0
dBm.
See Also Screens: RF Analyzer/RF Generator (RF Generator). Chapter 3:
Specifications
2. CW Freq,
CWFreqErr
This field selects the type of measurement to be made on the signal at the RF input.
This measurement is valid only for non-pulsed signals. The input frequency needs to
be within ±500 kHz of the value set as the Expected Input (4).
Choices CW Freq displays the measured frequency.
CWFreqErr displays the measured frequency error relative to the
continuous wave frequency set in the RF Analyzer’s Frequency
field.
See Also Keys: ON/OFF, HI LIMIT, LO LIMIT, REF SET, AVG,
METER, Units