4-29
Screens
Cell Control - Active Cell
13. Pwr Ramp This field displays the power mask and, (option 006 only), pulse measurements
screens.
14. Pwr Zero This field zeros the power meter. RF power must be disconnected from the RF
IN/OUT port when executing this function.
15. Reset This field resets all the Agilent 8922M/S’s measurements.
16. Scope This field displays the oscilloscope measurements screen.
17. Spec Anl (Option
006 only)
This field displays the spectrum analyzer measurements screen.
18. Mobile Reports The mobile reports reflect the status of the signal being received by the mobile
phones receiver.
19.TXLevel/Manual This field sets the expected input level of the RF Analyzer on the RF IN/OUT port.
This level must be within ±3 dB of the actual level in order that fully calibrated
measurements may be made.
This field displays the amplitude the Agilent 8922M/S is expecting from the mobile
phone, and adjusts its RF analyzer setting to match this input. When this field is
changed, the mobile phone TX Level and the Agilent 8922M/S RF Analyzer are no
longer coupled together. This means that you have independent control of the
amplitude of the transmitted mobile phone signal and the sensitivity of the
Agilent 8922M/S analyzer circuit.
When the fields are coupled, the Expected Input shows the power level of the
setting. When they are independent, the Expected Level displays Manual.
Range GSM900 and
E-GSM: −27.9 to +41.0 dBm.
The fields in the Mobile Phone area display information about
the call.
See Also Chapter 3: Specifications
20.Channel(Mobile) The channel field allows you to change the channel number of the call, either before
or during the call. If this field is changed when a call is in progress, a handover is
performed.