Agilent Technologies 8922M Cell Phone User Manual


 
3-30
Verifying Performance
Agilent Technologies 8922M/S Specifications
Reference
Specifications
The accuracy needs for testing GSM radios require the unit to be operated with the
High Stability Reference (Option 001) or an external high stability reference.
Fixed Reference Mode
Tunable Reference Mode
Allows offsetting the internal reference by a selected amount relative to the High
Stability Reference (Option 001) or an External Reference.
The Video Out connector on the rear panel outputs a 15 kHz PAL CVBS
underscanning compatible signal.
Accuracy (after warm up)
±[(Time since calibration × Aging rate) + Temperature Effects +
Accuracy of calibration]
External Reference Input
Frequency 13, 10, 5, 2, or 1 MHz,
±30 ppm
Level 0 to +10 dBm
Supplemental Characteristics
Nominal Impedance
50
10 MHz OUT (rear panel BNC)
Level
>+8.0 dBm nominal
Impedance
50 nominal
13 MHz OUT (rear panel BNC)
Level
>+8.0 dBm nominal
Impedance
50 nominal
Aging <2 ppm/year
Temperature Stability
±1 ppm (0° to 55°C)
Warm-up Time
<30 minutes,
±2ppm of final frequency
Required External Reference Accuracy
±0.5 ppm
Tune Range
±30 ppm
Reference Accuracy
±1 ppm + accuracy of External Reference or High Stability
(Option 001).
Temperature Stability
4 ppm, for selected offsets of up to ±30 ppm.