Agilent Technologies 8922M Cell Phone User Manual


 
8-30
Instrument BASIC
Programming and Using the TESTS Subsystem
Figure 8-12 Test Execution Conditions
To Have Testing Stop or Continue on a UUT Failure
Refer to item (1) in Figure 8-12 on page 8-30.
On UUT Failure.….….…. Continue
Testing continues whenever the UUT (Unit Under Test) fails to meet its test
specification limits. When this occurs, an error is listed on the test-results printout
and/or is displayed on the CRT.
On UUT Failure.….……Stop
Testing stops whenever the UUT fails to meet test specification limits.
To Continue or Pause After Each Test
Refer to item (2) in Figure 8-12 on page 8-30.
Run Mode.….……Continuous
All tests run in sequence. Testing pauses only if the operator is required to interact
with the UUT or Agilent 8922M/S; interaction such as changing UUT channels,
setting squelch, changing audio level, and so forth, cause testing to pause.
Run Mode.….……Single Step
The program stops running at the completion of each test. The test-system operator
is prompted to select Continue to proceed with testing.
2
4
1
3
5